Atlas FDA

Integra Licox Brain Tissue Oxygen Monitoring 1 x Complete Brain Probe Kit REF IM3.SR Rx Only, Manufacturer GMS- Gesellsc

FDA device recall Z-0058-2014 · posted 2013-10-23 · Terminated

Recall details

Recalling firm
Integra LifeSciences Corp.
Reason for recall
Integra LifeSciences Corporation has determined that there is a possibility that some lots of Licox CMP Brain Tissue Oxygen and Temperature Catheter Kits contain a hex wrench that does not loosen the set screw on the drill bit.
Action taken
Integra LifeSciences Corporation sent a recall letter/return response form on 6/11/2013 by traceable courier service, traceable e-mail or facsimile.
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
GWM
Quantity affected
175 KITS
Distribution
Worldwide distribution: US (nationwide) and countries of: AR, BR, CH, CL,CR. DE, ES, FI, FR, GB, IL, IT, MX, PA, PL, PT, RU, TW, and ZA.
Date initiated
2013-06-11
Date posted
2013-10-23
Date terminated
2014-02-18
Location
Plainsboro, NJ

Monitor recalls programmatically