Instrumentation Laboratory GEM Premier 4000 PN 00025000000 The GEM Premier 4000 is a portable critical care system for u

FDA device recall Z-2165-2011 · posted 2011-05-10 · Terminated

Recall details

Recalling firm
Instrumentation Laboratory Co.
Reason for recall
Measurement (amperometric) spike can occur early in cartridge life on the glucose and lactate sensors during patient blood analysis, leading to erroneously high results
Action taken
Instrumentation Laboratory notified accounts by a Field Safety NotifIcation Letter dated Feb 2011. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to use whichever of the following mitigation options works best for their institution. Both options will have the same effect to eliminate the potential impact of a rare spike on the glucose and lactate sensors. Customers were instructed to complete the enclosed Response Tracking Form and return immediately fax to 781-861-4207 or e-mail to ra-usa@ilww.com. Customers were advised to contact their local representative with any questions.
Root cause
Device Design
Status
Terminated
Product code
CGA
Quantity affected
4379 units
Distribution
Worldwide Distribution - USA (nationwide) and the countries of CANADA ISRAEL ITALY APAN KUWAIT MALAYSIA MEXICO NETHERLANDS NORWAY PANAMA POLAND PORTUGAL PUERTO RICO SAUDI ARABIA SLOVENIA SPAIN SWEDEN
Date initiated
2011-02-14
Date posted
2011-05-10
Date terminated
2012-05-23
Location
Bedford, MA