INSTRUMENT ARM DRAPE IS2000, 20 PACK: Product Usage: The intended use of the Instrument arm drape is to protect equipmen

FDA device recall Z-2101-2012 · posted 2012-07-27 · Terminated

Recall details

Recalling firm
Intuitive Surgical, Inc.
Reason for recall
Specific lots of the Instrument Arm Drapes were manufactured with a sterile adaptor that may have difficulty engaging an instrument.
Action taken
Intuitive Surgical sent an Urgent Medical Device Correction letters dated July 16, 2012 via Fed Ex to all affected customers in the US. The customers outside of the US were notified starting July 17, 2012. The letter identified the affected products, problem and actions to be taken. Customers were instructed to Identify affected lots, order replacement products and return both full and partial boxes of the affected products for credit as soon as unaffected products are available. All staff should be made aware of the issue. Customers were instructed to complete and return the acknowledgement form immediately. Questions and concerns should be directed to Intuitive Customer Service at 1-800-876-1310. EXPANSION: Intuitive discovered an additional lot (Lot DA121075) of Instrument Arm Drape IS2000, 20 pack that should have been included in the recall and expanded the recall on September 4, 2012. Letters to customers were sent by Fed Ex on September 5, 2012.
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
NAY
Quantity affected
92390 total instrument Drapes
Distribution
Worldwide Distribution - US Nationwide and the countries of: Australia, Spain, India, Italy and Saudi Arabia.
Date initiated
2012-07-13
Date posted
2012-07-27
Date terminated
2013-06-12
Location
Sunnyvale, CA