InPen smart insulin pen, REF: MMT-105NNBLNA, MMT-105NNGYNA, MMT-105NNPKNA, MMT-105ELBLNA, MMT-105ELGYNA, MMT105ELPKNA
FDA device recall Z-1650-2025 · posted 2025-04-21 · Open, Classified
Recall details
- Recalling firm
- Medtronic MiniMed, Inc.
- Reason for recall
- Insulin pens may have been incorrectly assembled therefore users could experience the following: either the insulin cartridge will not fit into the cartridge holder; or the cartridge holder may be difficult to remove from the insulin pen. If the issue is present, and if the user does not have backup insulin they may experience temporary hyperglycemia.
- Action taken
- On 3/4/2025, correction notices were sent to customers and distributors informing them of the following: 1) If you have already inserted your insulin cartridge into your device without difficulty removing the cartridge holder, then you can continue normal use. 2) If you have a device from one of the affected lot numbers and have not yet started using it, or have had difficulties in removing the cartridge holder or inserting the insulin cartridge, please call 1-800-646-4633, option 1, or go to https://info.medtronicdiabetes.com/ICH-replacement-form to request a replacement product. Do not use this device and dispose of it according to local regulations. 3) Complete and return the confirmation form via link: https://info.medtronicdiabetes.com/cartridge-holder Distributors are asked to notify impacted customers. If you have further questions or need assistance, please call the firm's 24-Hour Technical Support line at 1-800-646-4633, option 1.
- Root cause
- Mixed-up of materials/components
- Status
- Open, Classified
- Product code
- FMF
- Quantity affected
- 8,986
- Distribution
- US: OK, IA, NC, SC, MI, MN, CA, AZ, AL, MO, WA, IL, KS, TX, FL, MT, MD, OH, VA, NY, TN, HI, MA, NJ, PA, CO, WI, GA, CT, NE, OR, LA, NM, SD, ND, IN, NV, WV, RI, AR, NH
- Date initiated
- 2025-03-04
- Date posted
- 2025-04-21
- Location
- Northridge, CA
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| InPen System App (MMT-8060 (iOS), MMT-8061 (Android)) (K242775) | Medtronic Minimed | 2024-11-12 |
| InPen System (K201337) | Companion Medical, Inc. | 2020-06-18 |
| InPen Dose Calculator (K190487) | Companion Medical, Inc. | 2020-02-18 |
| InPen System (K192841) | Companion Medical, Inc. | 2020-02-10 |
| InPen (K160629) | Companion Medical, Inc. | 2016-07-26 |