Atlas FDA

INOvent, Nitric Oxide Delivery System. Model Number 1605-9000-000

FDA device recall Z-1410-06 · posted 2006-08-25 · Terminated

Recall details

Recalling firm
INO Therapeutics, Inc.
Reason for recall
The Kel-F Tip mounted on the high pressure hose can become dislodge/lodged in the INOmax valve outlet.
Action taken
On June 29, 2006, the firm sent a letter to its consignees advising them to visually verify that the Kel F tip is in place before attempting to connect either of the high pressure hoses to the INOmax cylinders. INO advised that if the tip was not in place, the operator should immediately replace the tip before attaching the hoses. On July 5, 2005, the firm sent a second letter to their consignees with extra Kel F tips and existing instructions from the INOvent Operation & Maintenance Manual outlining the proper replacement procedure for the tips. The firm is following with telephone contact to assure that the letters were delivered to the proper person(s).
Root cause
Other
Status
Terminated
Product code
MRN
Quantity affected
3,123 units
Distribution
Worldwide, USA and countries of Canada, Uruguay, Chile, Argentina, Colombia, and Malaysia
Date initiated
2006-06-29
Date posted
2006-08-25
Date terminated
2007-07-16
Location
Clinton, NJ

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
DATEX-OHMEDA INOVENT DELIVERY SYSTEM (K000186)Datex-Ohmeda2000-02-10