Innova 4100IQ Indicated for use in generating fluoroscopic images of human anatomy for vascular angiography, diagnostic
FDA device recall Z-0453-2015 · posted 2014-12-16 · Terminated
Recall details
- Recalling firm
- GE Healthcare, LLC
- Reason for recall
- Loss of imaging - affected devices may experience a loss of real-time interventional imaging and/or gantry motion due to the Electromagnetic (EM) sensitivity of the grid detection mechanism.
- Action taken
- Consignees were sent on 9/4/2014 a GE Healthcare "Urgent Medical Device Correction" letter GEHC Ref#12218 dated September 4, 2014. The letter was addressed to Director of Biomedical Engineering, Director of Radiology and Chief of Cardiology. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction, and Contact Information. Customers are to perform a system functional check as instructed in the letter. If the check fails, they should contact their local GE Healthcare Service Representative. GE Healthcare will correct all systems at no cost to the customer, and a representative will contact customers to arrange for the correction. For questions they can call US 800-437-1171. For other countries they can call their local GE Healthcare Service Representative.
- Root cause
- Device Design
- Status
- Terminated
- Product code
- OWB
- Quantity affected
- 6 (1 US, 5 OUS)
- Distribution
- Worldwide Distribution -- US, including the states of CA, CO, FL, MO, PA; and, the countries of BELGIUM, CHINA, COLOMBIA, FRANCE, GREECE, MEXICO, and RUSSIA.
- Date initiated
- 2014-09-04
- Date posted
- 2014-12-16
- Date terminated
- 2015-01-21
- Location
- Waukesha, WI