Atlas FDA

Innova 3100 / 3100 IQ, Cardiovascular Imaging System, GE Healthcare, Waukesha, WI 53188 (The system consists of an a mon

FDA device recall Z-2437-2008 · posted 2008-09-20 · Terminated

Recall details

Recalling firm
GE Healthcare
Reason for recall
The charger and / or battery of the PDB may fail earlier than expected with no advance warning and thus cause the system to shut down.
Action taken
Consignees (Hospital Administrators; Hospital Risk Management Department; Mangers of Radiology/Cardiology; Radiologists/Cardiologists) were sent a "GE Urgent Medical Device Correction" letter dated July 2008. The letter described the affected products; the problem; and the user instructions, which state that the hospital should establish emergency workflow procedures to ensure patient safety should a component stop operating during a patient examination. Facilities should verify if their system has been impacted by checking if the catalog numbers listed are referenced on the PDB panel located in the equipment room. The letter also states the GE Healthcare Resolution Plan, includes correcting the condition, and contact Information.
Root cause
Device Design
Status
Terminated
Product code
MQB
Quantity affected
375
Distribution
Nationwide Distribution --- including states of CA, FL, NY, AK, AL, AR, AZ, PA, FL ,CO, SC, CT, IA, DC, DE, GA, HI, IL , IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, NJ, MS, MT, NC, ND, NE, NH,
Date initiated
2008-07-25
Date posted
2008-09-20
Date terminated
2011-12-10
Location
Waukesha, WI

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
INNOVA 2100-IQ, INNOVA 3100, INNOVA 3100-IQ, INNOVA 4100, AND INNOVA 4100-IQ (K052412)GE Healthcare2005-09-16
INNOVA CT OPTION FOR DIGITAL FLUOROSCOPIC IMAGING SYSTEMS INNOVA MODELS 4100, 4100IQ, 3100, 3100IQ & 2100IQ (K052157)Ge Medical Systems, LLC2005-08-25
INNOVA 4100, INNOVA 3100 (K042053)Ge Medical Systems, Inc.2004-08-26
INNOVA 3100 (K031637)GE Medical Systems2003-08-22