Ingenuity TF PET/MR, Diagnostic Imaging System , Philips Medical Systems

FDA device recall Z-2491-2015 · posted 2015-08-31 · Terminated

Recall details

Recalling firm
Philips Medical Systems (Cleveland) Inc
Reason for recall
Philips Healthcare has become aware of a problem in which the Ingenuity TF PET/MR PET Reconstruction Server (PRS) database may lock up after an MR acquisition is completed and before a PET acquisition is begun, even though no actual database lockup issue has been reported in the field-installed Ingenuity TF PET/MR systems.
Action taken
On July 23, 2015 the firm sent Field Safety Notices to their customers.
Root cause
Software design
Status
Terminated
Product code
OUO
Quantity affected
11
Distribution
Product was shipped to the following states: NY, OH & TX. Product was also shipped to the following countries: Finland, Germany, Hong Kong, Japan, Netherlands, South Korea, Spain & Switzerland.
Date initiated
2015-07-23
Date posted
2015-08-31
Date terminated
2018-06-26
Location
Cleveland, OH