Ingenia Elition S Magnetic Resonance (MR) system, Product Numbers 781357 and 782106

FDA device recall Z-2396-2023 · posted 2023-08-11 · Open, Classified

Recall details

Recalling firm
Philips North America
Reason for recall
The system operator is able to bypass the SmokeDetector Interlock system locking mechanism after a smoke detection by power cycling the system.
Action taken
An URGENT Medical Device Correction Notice dated 6/5/23 was sent to customers. Actions that should be taken by the customer / user in order to prevent risks for patients or users: 4.1. Philips has installed a SmokeDetector Interlock feature on all the systems listed in Appendix A. Post this notice near the affected MR system(s) for ease of reference. Circulate this notice to all users of this device so that they are aware of the product issue and associated hazard/harm until this issue has been resolved. Please complete and return the attached updated customer response form to Philips promptly and no later than 30 days from receipt of this letter. Philips is requesting that even if you acknowledged a previous version of this letter that you also respond with an acknowledgment to this updated letter. 4.2. The SmokeDetector Interlock feature is integrated with the MR System and powers off the relevant part of the system (Gradient Amplifier) when smoke is detected in order to prevent further development of smoke and/or fire. If the SmokeDetector Interlock has detected smoke and has powered off the Gradient Amplifier: Remove patient from the system according to the instructions for use because power to the tabletop will still be active Scanning is disabled until Philips service has checked the system. Do not attempt to continue scanning. Immediately inform Philips Service DO NOT attempt to restart or perform a power cycle of the system 4.3. If a user experiences smoke, fire, or severe image quality issues (image artifacts caused by electrical disturbances): Immediately stop scanning and evacuate the patient from the Examination Room. Check Examination Room for a developing fire. If a developing fire is sensed, adhere to established fire emergency procedures, which may include switching off power to the complete system and/or removing the magnet field by using the Emergency Magnet Off button Scanning is disabled until Philips service has checked t
Root cause
Under Investigation by firm
Status
Open, Classified
Product code
LNH
Quantity affected
640 total
Distribution
US Nationwide. Global Distribution.
Date initiated
2023-06-12
Date posted
2023-08-11
Location
Cambridge, MA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
Ingenia 3.0T, Ingenia 3.0T CX, Ingenia Elition and MR 7700 with Distributed Multi Nudclei (K213516)Philips Medical Systems Nederland B.V.2022-03-03
Achieva, Intera, Ingenia, Ingneia CX, Ingenia Elition, and Ingenia Ambition MR Systems (K193215)Philips Medical Systems Nederland B.V.2020-04-10
Ingenia, Ingenia CX, Ingenia Elition, and Ingenia Ambition MR Systems (K183063)Philips Medical Systems Nederland B.V.2019-02-14
Ingenia Elition S and Ingenia Elition X (K173451)Philips Medical Systems Nederland B.V.2018-03-20
Ingenia Elition S and Ingenia Elition X R5.4 (K173356)Philips Medical Systems Nederland B.V.2018-03-06