Ingenia Ambition S - Model Number: 781359 Diagnostic MR system.
FDA device recall Z-1796-2021 · posted 2021-06-04 · Open, Classified
Recall details
- Recalling firm
- Philips North America Llc
- Reason for recall
- The MR magnet is equipped with 4 lifting blocks, which are positioned on top of the magnet. These blocks are used to allow the magnets to be moved in or out by a lifting crane, e.g. during decommissioning or moving activities. A mechanical issue was detected on these hoisting interface blocks on top of the magnet.
- Action taken
- On March 23, 2021, Philips issued an Urgent Medical Device Correction notice to customers via letter stating Philips identified an issue with specified MR systems. The MR magnet is equipped with 4 lifting blocks, which are positioned on top of the magnet. These blocks are used to allow the magnets to be moved in or out by a lifting crane, e.g. during decommissioning or moving activities A mechanical issue was detected on these hoisting interface blocks on top of the magnet. In the initial design of the hoisting blocks, the diameter to meet the required bolt-on interface surface for the swivel eye bolts, used as lifting hardware, did not meet the required specification. HOW TO IDENTIFY AFFECTED PRODUCTS Ingenia Ambition X (781356) and Ingenia Ambition S (781359) ACTION TO BE TAKEN BY CUSTOMER / USER No action is required by the customer, as the clinical use of the MR system is not affected. In case of a planned decommissioning or movement of the magnet, and this FCO is not installed yet, please contact your local Philips representative. ACTIONS PLANNED BY PHILIPS Via FCO781 00524 a mechanical solution will be rolled out to the affected MR systems. ***Updated November 2023*** Additional customers were identified and notified on August 22, 2023.
- Root cause
- Device Design
- Status
- Open, Classified
- Product code
- LNH
- Quantity affected
- 156 MR systems
- Distribution
- Domestic: FL, GA, MI, MO, NH, NJ, NY, PA, SC, TX, VT. Foreign: Australia, Austria, Bangladesh, Belgium, Brazil, Denmark, Finland, France, Germany, Iraq, Italy, Japan, Libya, Moldova, Netherlands, Oman
- Date initiated
- 2021-04-23
- Date posted
- 2021-06-04
- Location
- Cambridge, MA