Infusion Dynamics Power Infuser pump. Model number M100B-3A

FDA device recall Z-0062-05 · posted 2004-11-02 · Terminated

Recall details

Recalling firm
Infusion Dynamics, A Div. of Zoll Med Corpinfusion Dynamics
Reason for recall
Front Panel may be cracked on some units. If cracked fluid can enter the infuser and short the start/stop button. Shorting would cause the infuser to stay in the open position.
Action taken
The recalling firm issued a recall letter dated 9/20/04 to their customers informing them of the problem and the need to return the units.
Root cause
Other
Status
Terminated
Product code
FRN
Quantity affected
255 units
Distribution
The products were shipped to dealers in TX, MO, and VT; medical facilities in CA, NV, and NY. The product was also shipped to government accounts in PA and Germany.
Date initiated
2004-09-21
Date posted
2004-11-02
Date terminated
2005-02-07
Location
Plymouth Meeting, PA