Atlas FDA

Infinity Acute Care System (IACS) Monitoring Solution: The IACS is intended for multi-parameter, physiologic patient mon

FDA device recall Z-1104-2021 · posted 2021-02-10 · Open, Classified

Recall details

Recalling firm
Draegar Medical Systems, Inc.
Reason for recall
The Infinity M540 patient monitor may randomly reboot due to an error to correctly transmit and read the header data of files in the memory of the device. Under this situation, the device will try to reboot to mitigate the error. The device will be available again for use within 30 seconds. If this error continues and the M540 reboots three times in a time span of 10 minutes, it will enter a fail-state. A fail-state will annunciate itself with a continuous sound to alert the user. The M540 will reset to factory default and the user can manually configure and readmit the patient to continue patient monitoring.
Action taken
On 01/2021, Draeger issued Urgent Medical Device Correction notices via certified mail. The purpose of the letter is to advise customers that Draeger is voluntarily recalling our Infinity Acute Care System and the Standalone Infinity M540 patient monitors due to a software issue.
Root cause
Software design
Status
Open, Classified
Product code
MHX
Quantity affected
19802 systems (Domestic: 3412, Foreign 16,390)
Distribution
Worldwide distribution - US Nationwide distribution including in the states of AL, AZ, CA, CO, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, ME, MI, MN, MO, MS, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, S
Date initiated
2020-12-23
Date posted
2021-02-10
Location
Andover, MA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
INFINITY ACUTE CARE SYSTEM, INFINITY MCABLE- MASIMO RAINBOW SET (K113798)Draeger Medical Systems, Inc.2012-03-22
INFINITY ACUTE CARE SYSTEM (IACS) MONITORING SOLUTION (K093788)Draeger Medical Systems, Inc.2010-03-08