Infinite 200 The Tecan Infinite 200 is a multifunctional microplate reader with injector option. The Infinite 200 provid

FDA device recall Z-0991-2015 · posted 2015-01-15 · Terminated

Recall details

Recalling firm
Tecan US, Inc.
Reason for recall
Incorrect lumi firmware version installed (E.027 instead of V2.00)
Action taken
Tecan sent an Urgent Field Corrective Action letter dated October2014, to all affected consignees by via FedEx on/or about November 10, 2014. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed not to use te kinetic luminescence measurement workflow requiring movement of th eplate transport out during kinectice measurements until the firmware has been upgraded to current version 2.0. A Tecan service representative will contact you to arrange a time to install the firmware. Additionally, they may want to review prior measurement results obtained with this workflow. Customers with questions were advised to contacte their local Tecan Helpdesk. For questions regarding this recall call 919-361-5200, ext 19524.
Root cause
Employee error
Status
Terminated
Product code
KHO
Quantity affected
8 units
Distribution
Worldwide Distribution - USA including MD and Internationally to Hong Kong, South Korea and the UK.
Date initiated
2014-11-10
Date posted
2015-01-15
Date terminated
2015-12-21
Location
Morrisville, NC