Infinite 200 The Tecan Infinite 200 is a multifunctional microplate reader with injector option. The Infinite 200 provid
FDA device recall Z-0991-2015 · posted 2015-01-15 · Terminated
Recall details
- Recalling firm
- Tecan US, Inc.
- Reason for recall
- Incorrect lumi firmware version installed (E.027 instead of V2.00)
- Action taken
- Tecan sent an Urgent Field Corrective Action letter dated October2014, to all affected consignees by via FedEx on/or about November 10, 2014. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed not to use te kinetic luminescence measurement workflow requiring movement of th eplate transport out during kinectice measurements until the firmware has been upgraded to current version 2.0. A Tecan service representative will contact you to arrange a time to install the firmware. Additionally, they may want to review prior measurement results obtained with this workflow. Customers with questions were advised to contacte their local Tecan Helpdesk. For questions regarding this recall call 919-361-5200, ext 19524.
- Root cause
- Employee error
- Status
- Terminated
- Product code
- KHO
- Quantity affected
- 8 units
- Distribution
- Worldwide Distribution - USA including MD and Internationally to Hong Kong, South Korea and the UK.
- Date initiated
- 2014-11-10
- Date posted
- 2015-01-15
- Date terminated
- 2015-12-21
- Location
- Morrisville, NC