Atlas FDA

Infinite 200 in vitro diagnostic. Product Usage: The Tecan Infinite 200 is a multifunctional microplate reader with inje

FDA device recall Z-1971-2015 · posted 2015-07-02 · Terminated

Recall details

Recalling firm
Tecan US, Inc.
Reason for recall
A firmware bug causes incorrect values to be displayed for specific measured luminescence values.
Action taken
Tecan sent an Urgent Field Corrective Action letter dated May 29, 2015 to their customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to complete the attached Customer Response Form. For questions contact your local Tecan Helpdesk.
Root cause
Device Design
Status
Terminated
Product code
KHO
Quantity affected
868 in total
Distribution
US Nationwide Distribution in the states of AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MO, NE NH, NJ, NM, NY, NC, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, W
Date initiated
2015-05-29
Date posted
2015-07-02
Date terminated
2017-08-18
Location
Morrisville, NC