Infant Ventilators with Electronic Manometer. Models IV-100B; IV-200 and IV-200 SAVI.

FDA device recall Z-1032-04 · posted 2004-07-20 · Terminated

Recall details

Recalling firm
Sechrist Industries Inc
Reason for recall
Electronic manometer malfunctions. Manometer may be illegible. The alarm for power failure may function only briefly, or not at all.
Action taken
Recall notification contains a test method to be used to identify defective units. Return notification that ventilator is defective will result in replacement of the electronic manometer.
Root cause
Other
Status
Terminated
Product code
CBK
Quantity affected
240
Distribution
Nationwide and to Malaysia, Japan, Kuwait, Sri Lanka, South Africa, Guatemala, Mexico, Morocco, Indonesia, Korea, Ecuador, Colombia, Portugal, Oman, Switzerland, Spain, Zimbabwe, Belgium, Brazil, Thai
Date initiated
2004-05-20
Date posted
2004-07-20
Date terminated
2006-03-14
Location
Anaheim, CA