Atlas FDA

in2it (I) and (II) System A1c Test cartridges and in2it Self Test System A1c Test Cartridges, for use with the in2it (I)

FDA device recall Z-0122-2011 · posted 2010-10-22 · Terminated

Recall details

Recalling firm
Bio-Rad Laboratories Inc
Reason for recall
Inaccurate results with high bias-- Inaccurate results may be read by the machine, creating the risk for untreated hypoglycemia at elevated altitudes.
Action taken
BIO-RAD Laboratories Deeside Ltd., sent an URGENT: MEDICAL DEVICE CORRECTION letter, dated July 26, 2010, to their subsidiaries and to all customers identifying the product, the problem, and the action to be taken. Subsidiaries were to ensure that all customers received the letter with the Customer Medical Device Correction Response Form and collect the completed form from every customer. Once forms were received from all customers they were to be faxed to BIO-RAD Deeside Regulatory Affairs at +44 (0) 1244833401. Customers were instructed to run a test and visually inspect the test cartridge to detect any residual clear liquid. If the cartridge did not run properly the test was to be rejected. Customers were to run Quality Control test with each new shipment of test cartridges. Patient's results were not to be reported if the QC failed. Customers were to complete the CUSTOMER MEDICAL DEVICE CORRECTION RESPONSE FORM and fax it to the number provided on the form. For questions customers were to contact their regional BIO-RAD office.
Root cause
Device Design
Status
Terminated
Product code
LCP
Quantity affected
90703
Distribution
Product was distributed to Bio-Rad's US subsidiary, along with 18 other distribution subsidiaries outside the US. These distributed further to other consignees. POroduct was distributed to Mexico, Kor
Date initiated
2010-07-26
Date posted
2010-10-22
Date terminated
2011-10-25
Location
Hercules, CA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
G5 I HBA1C TEST, G5 II HBA1C TEST (K041635)Provalis Diagnostics , Ltd.2004-08-16