IMx Sirolimus Reagent Pack; list 5C91-21; 100 test pack containing 1 bottle (9.7 mL) Anti-Sirolimus Antibody Coated Micr

FDA device recall Z-0501-2007 · posted 2007-02-15 · Terminated

Recall details

Recalling firm
Abbott Laboratories MPG
Reason for recall
The sirolimus values may shift higher after storage at 2-8C or after one freeze/thaw cycle of specimens.
Action taken
Abbott initiated a field correction recall by letter dated 8/11/06 to all customers receiving the affected reagent. The accounts were informed of the shift in sample values after storage at 2-8*C or with freeze thaw of specimens with the affected lots of reagents. The accounts were requested to review the revised sample handling information included in the attachment to the letter with their laboratory personnel and retain a copy of the letter and attachment for their records. Customers requiring additional information were directed to contact Abbott Customer Service at 1-877-422-2688.
Root cause
Other
Status
Terminated
Product code
NRP
Quantity affected
2,800 kits
Distribution
Worldwide, including USA, Puerto Rico, Poland, Croatia, Latvia, Estonia, Serbia & Montenegro, Japan, Turkey, Singapore, China, India, Taiwan, Thailand, South Africa, South Korea, Australia, New Zealan
Date initiated
2006-08-11
Date posted
2007-02-15
Date terminated
2008-01-18
Location
Abbott Park, IL