Implant Epikut S Plus CM 16-Degrees 4,5 X 18,0 mm, REF: ILM 4518N

FDA device recall Z-0920-2025 · posted 2025-01-15 · Open, Classified

Recall details

Recalling firm
SIN SISTEMA DE IMPLANTE NACIONAL S.A
Reason for recall
Dental implant box incorrectly labels the diameter as 4,5 X 18,0 mm, but the internal label correctly labels the diameter as 4,0 X 18,0 mm; if the dentist drills the bone with the larger diameter stated on the box, the smaller diameter implant, will become loose and will not fit, which may necessitate an additional surgery.
Action taken
On 9/1/2024, "URGENT: MEDICAL DEVICE RECALL" notices were emailed to customers and distributors who were told the following: The product should not be used. It does not pose a handling risk and should be separated and discarded. Complete and return the Medical Device Recall Return Response Acknowledgement and Receipt Form via email to: brian@sindentalusa.com. Customer with questions may contact the company via telephone at 0800-770-8290 between the hours of 9 and 19 (Brasilia time) or email the firm at anna.lopes@sinimplantsystem.com.
Root cause
Process control
Status
Open, Classified
Product code
DZE
Quantity affected
123 (US 98, OUS 25)
Distribution
US distribution to states of: KY, CO, CA, NY, ID, NJ, FL, TX, KY, IL, OH; AND OUS (foreign) country of Brazil.
Date initiated
2024-09-01
Date posted
2025-01-15
Location
Sao Paulo, Brazil