IMPAX Agility. IMPAX Agility when deployed as a departmental Picture Archiving and communications system (PACS) provides
FDA device recall Z-2061-2015 · posted 2015-07-13 · Terminated
Recall details
- Recalling firm
- AGFA Healthcare Corp.
- Reason for recall
- Firm received a complaint that the hospital had sent two accession numbers over the same morning and on the Agility side the patient name was incorrect.
- Action taken
- PRB0046372 was created to address the corrective actions related to this issue. On 6/26/2015, an email communication, as an effectiveness check was sent to the two customers. Acknowledgment, via email, that the communication was received was requested from the consignees.
- Root cause
- Other
- Status
- Terminated
- Product code
- LLZ
- Quantity affected
- 2
- Distribution
- Distributed in the state of MO.
- Date initiated
- 2015-06-26
- Date posted
- 2015-07-13
- Date terminated
- 2016-01-27
- Location
- Greenville, SC