IMO-402 Mononucleosis Rapid Test Device (Whole Blood/Serum/Plasma). Catalog #IMO-402/94201 The MONO test (Whole Blood) i

FDA device recall Z-1148-2013 · posted 2013-04-18 · Terminated

Recall details

Recalling firm
Alere San Diego, Inc.
Reason for recall
Alere San Diego is recalling the Clearview Mononucleosis Cassette because the positive control included in the rapid test may produce false negative results.
Action taken
Alere sent an Urgent Medical Device Recall letter dated February 26, 2013, to all affected customers. The letter informs the customers of the problems identified and the actions to be taken. Customers with unused product remaining are instructed to immediately discard it. Investigations of customer complaints indicate that the positive control included with this lot may produce false negative results. Customers are instructed to return the attached Verification form. Customers that have consumed all product are instructed to indicate zero under "Quantity Discarded" and complete and fax back the form. Replacement product will be provided for any unused portion of the affected lot based on completion and return of the attached Verification Form. Customers with any questions are instructed to contact Alere Technical Services at (866) 216-0094 or by email at later.flow.support@alere.com. For questions regarding this recall call 858-805-2000, ext 3015.
Root cause
Material/Component Contamination
Status
Terminated
Product code
KTN
Quantity affected
78 kits
Distribution
Nationwide Distribution including TN, WA, UT, KS, KY, MN, GA, MA, PA, NY, VT, and the Dist. of Columbia
Date initiated
2013-02-26
Date posted
2013-04-18
Date terminated
2013-08-29
Location
San Diego, CA