ImmunoCAP Allergen Ri206, Cockroach American, Periplaneta americana It is an analyte specific reagent and no intended us

FDA device recall Z-1055-2012 · posted 2012-02-17 · Terminated

Recall details

Recalling firm
Phadia US Inc
Reason for recall
The firm found this lot of ImmunoCAP Allergen Ri206, American Cockroach does not contain sufficient levels of some cockroach allergens e.g. tropomyosin. Approximately 2% of the positive samples have a risk to fall below the Limit of Quantization (o .1 kUA/L).
Action taken
ThermoFisher sent an URGENT RECALL NOTIFICATION letter dated December 22, 2011, to all affected consignees. The letter identifed the product, the problem, and the action to be taken by the conignee. Consignees were recommended to cease testing on the affected lot of product and, if so desired, substitute with ImmunoCAP Allergen i6, German Cockroach, until ImmunoCAP Allergen Ri206 was in stock. For any questions regarding this notification, contact Technical Support at 800.346.4364, option 2.
Root cause
Release of Material/Component prior to receiving test results
Status
Terminated
Product code
MVU
Quantity affected
780; all sent to US, with 10 assays per sleeve
Distribution
Nationwide to include: MN, VA, NC, MO, SC, TX, MA, MI, FL, IL, PA, MD, UT, CO, GA, CA,
Date initiated
2012-01-03
Date posted
2012-02-17
Date terminated
2013-02-04
Location
Portage, MI