Immulite 2000. Immunoassay system. Software version 2.6 was to be uploaded on systems running 2.5b software.

FDA device recall Z-1138-03 · posted 2003-08-20 · Terminated

Recall details

Recalling firm
DPC Cirrus
Reason for recall
Version 2.6 software was released however it will not properly handle adjustment slopes.
Action taken
Recall Technical Bulletins were sent out on 6/13/2003 to all Immulite 2000 users instructing them to not install the provided version 2.6 software.
Root cause
Other
Status
Terminated
Product code
JJE
Quantity affected
434 upgrades
Distribution
Software was distributed to Immulite 2000 users throughout the nation. These units are found in hospitals and clinics. Software was also distributed to the firm''s service representatives.
Date initiated
2003-06-13
Date posted
2003-08-20
Date terminated
2004-11-01
Location
Flanders, NJ