IMMAGE Immunochemistry Systems Rheumatoid Factor (RF) Reagent, PN 447070 The IMMAGE Immunochemistry Systems Rheumatoid F

FDA device recall Z-0003-2013 · posted 2012-10-01 · Terminated

Recall details

Recalling firm
Beckman Coulter Inc.
Reason for recall
The recall was initiated because Beckman Coulter has confirmed customer reports of QC shifts, lot-to-lot variation, and an increase in the percent of patient samples recovering >20 IU/mL using the Lots# M012376 and M101865 of IMMAGE RF reagent. The lots identified for the RF reagent might produce false abnormal results.
Action taken
The firm, Beckman Coulter, sent an "URGENT PRODUCT CORRECTION" letter dated January 27, 2012 with an attached PCA Response Form to all customers who purchased the IMMAGE Immunochemistry Systems Rheumatoid Factor (RF) Reagent. The letter describes the product, problem and actions to be taken. The customers were instructed to discontinue use of IMMAGE RF reagent lots M012376 and M101865; call Order Assistance at (800) 526-3821 in the United States, (800) 463-7828 in Canada or contact your local Beckman Coulter representative for replacements; and complete and return the enclosed FAX BACK RESPONSE FORM within 10 days via mail at: Beckman Coulter, Inc., 250 S. Kraemer Blvd, Mail Stop A2.0246, Brea, CA 92822-8000; Attn: Regulatory Affairs or fax to: (714) 961-4234. Note: Beckman Coulter has added additional release testing that will assure future lots RF will meet the reference interval claim of the IMMAGE RF Instructions for Use (IFU). If you have any questions concerning this letter, contact Beckman Coulter Support Center (call Center) on the Beckman Coulter Website at: http://www.beckmancoulter.com/customersupport/support/ or call (800) 854-3633 in the US and Canada. Outside of the US and Canada, customers were instructed to contact their local Beckman Coulter Representative.
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
DHR
Quantity affected
925 units
Distribution
Worldwide distribution: USA (nationwide) and country of: Canada.
Date initiated
2012-01-27
Date posted
2012-10-01
Date terminated
2013-05-30
Location
Brea, CA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
IMMAGE SYSTEM RHEUMATOID FACTOR (RF) REAGENT (K963048)Beckman Instruments, Inc.1996-11-04