IMAGEnet 6 v1.53- IMAGEnet 6 Ophthalmic Data System is a software program that is intended for use in the collection, st

FDA device recall Z-1138-2019 · posted 2019-04-12 · Terminated

Recall details

Recalling firm
Topcon Medical Systems, Inc.
Reason for recall
When a patient is selected in the patient data and another patient is selected prior to the display of the primary patient's thumbnails is completed, then the primary patient's thumbnails are displayed with secondary patient's identification information such as name, patient's ID, etc. on the monitor screen of the PC used.
Action taken
Topcon issued Device Correction Notice on 3/22/19 providing problem- there exists a situation in which the displayed patient name and the thumbnail image do not match and action to take.Topcon has corrected this issue in IMAGEnet 6 version 2.0.3 and will be providing a remote upgrade free of charge. Contact customer support team to schedule the upgrade at 866.922.6278 option 1 or by email at TMSservice@topcon.com
Root cause
Labeling design
Status
Terminated
Product code
NFJ
Quantity affected
450 units
Distribution
Nationwide
Date initiated
2019-03-06
Date posted
2019-04-12
Date terminated
2019-07-17
Location
Oakland, NJ