Illumina Worklist Manager (IWM) (software v1.0.15), a component of Illumina MiSeqDx Platform. Product Usage: The Illumin

FDA device recall Z-2611-2014 · posted 2014-09-09 · Terminated

Recall details

Recalling firm
Illumina Inc
Reason for recall
A software component of the MiSeqDx instrument, called the Illumina Worklist Manager (IWM) v1.0.15, fails to perform as intended
Action taken
Illumina notified affected customers via phone, email and Product Quality Notification letter dated July 28, 2014 to describe the software anomaly. The letter identified the affected product, problem and actions to be taken. For questions contact your local support team or Illumina Technical Support at techsupport@illumina.com.
Root cause
Device Design
Status
Terminated
Product code
PFF
Quantity affected
35
Distribution
Worldwide Distribution - US (nationwide) and the countries of: France, Germany, Italy, Norway, UK and United Arab Emirates.
Date initiated
2014-07-28
Date posted
2014-09-09
Date terminated
2014-11-12
Location
San Diego, CA