iLab Ultrasound Imaging System; iLab 3.0 (Polaris 1.0) software. Used for ultrasound examinations of intravascular patho

FDA device recall Z-1487-2014 · posted 2014-04-22 · Terminated

Recall details

Recalling firm
Boston Scientific Corporation
Reason for recall
Specific iLab Ultrasound Imaging Systems using newest 3.0 (Polaris 1.0) software change the pullback speed from the default setting of 0.5 mm/sec to 1.0 mm/sec, the Longview distance measurement may display an inaccurate value that is double the actual distance.
Action taken
Boston Scientific sent an Urgent Medical Device Field Correction letter to all affected customers via hand delivery by BSC sales representatives. The letter identifed the affected product, problem and actions to be taken. Customers were asked to immediately post the Field Correction information in a visible location near the system(s) to ensure it is easily accessible to all users of the device. For question contact your local Sales Representative.
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
IYN
Quantity affected
9 units
Distribution
USA Nationwide Distribution in the states of NY, NC, DC, FL, PA
Date initiated
2014-04-03
Date posted
2014-04-22
Date terminated
2015-03-18
Location
Fremont, CA