IdentiTi ALIF Standalone Interbody System: Part Number/Description 132-10-85-250 / IdentiTi ALIF SA Graft Bolt, 08.5 x 2
FDA device recall Z-1182-2024 · posted 2024-02-23 · Open, Classified
Recall details
- Recalling firm
- Alphatec Spine, Inc.
- Reason for recall
- Due to reports of intraoperative graft bolt implantation breakages.
- Action taken
- On 01/26/2024, the firm emailed an "URGENT-MEDICAL DEVICE RECALL NOTIFICATION" Letter to customer informing them that, The Surgical Technique Guide, LIT-85141, and Instructions for Use, INS-130, were revised to require hole preparation prior to insertion of an IdentiTi SA Graft Bolt, as well as to add a new caution statement, to confirm with fluoroscopy that the graft bolt insertion angle is as close as possible to a 40-degree trajectory. The revised labeling mitigates potential fracturing of the Graft Bolt intra-operatively or post-operatively. Customer are instructed to: Please share this notice with all of those who need to be aware within your organization or any organization where the affected product has been transferred. - Inform doctors and surgical staff of the labeling (surgical technique and instructions for use) - Please fill out the last page of this letter to confirm that you have read this notification and will complete the necessary actions as described in this notification. - If you have any questions/concerns or require assistance with the correction, please contact by e-mailing QMSCompliance@atecspine.com or calling 1-800-922-1356.
- Root cause
- Device Design
- Status
- Open, Classified
- Product code
- OVD
- Quantity affected
- 12,777 devices
- Distribution
- Worldwide - U.S. Nationwide distribution including in the states of AL, AZ, CA, CO, CT, FL, GA, IA, ID, IN, KS, KY, MD, MI, MN, MS, NC, NH, NV, NY, OH, OK, OR, PA, Puerto Rico, SC, SD, TN, TX, UT, WA
- Date initiated
- 2024-01-26
- Date posted
- 2024-02-23
- Location
- Carlsbad, CA
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| IdentiTi ALIF Interbody Systems (K232097) | Alphatec Spine, Inc. | 2023-09-25 |