Icelock Expulsion Valve 551 Model L-551000 The valve has a push button for suction release to release a lower limb prost
FDA device recall Z-2026-2009 · posted 2009-09-04 · Terminated
Recall details
- Recalling firm
- Ossur Engineering, Inc
- Reason for recall
- The plastic caps have come off or become loose when the user stepped into their prosthesis.
- Action taken
- Customers were sent a Safety Notice Voluntary Product Recall, dated June 23, 2009, that advised them to return any unused valves to Ossur for a free replacement. Those using valves in combination with a cushion liner and suspension sleeve should replace immediately if problems are experienced, otherwise replace at next routine service. Customers using valve in combination with an Iceross Seal-In liner should replace ASAP and return used valves to Ossur for free replacement. Customers with questions can call Darlyne Roa 1-800-233-6263 ext 1023743.
- Root cause
- Process change control
- Status
- Terminated
- Product code
- ISP
- Quantity affected
- 4,647
- Distribution
- Worldwide Distribution
- Date initiated
- 2009-06-23
- Date posted
- 2009-09-04
- Date terminated
- 2012-03-06
- Location
- Albion, MI
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.