I-STAT EG7+ Cartridge Abbott Point of Care, Princeton, NJ. The i-Stat EG7 cartridge is intended to be used for quantitat

FDA device recall Z-2133-2012 · posted 2012-08-02 · Terminated

Recall details

Recalling firm
Abbott Point Of Care Inc.
Reason for recall
i-Stat EG7 cartridges may generate falsely depressed PCO2 results for samples above 26 mm hg when stored at room temperature for four weeks or longer.
Action taken
Abbott Point of Care, Inc sent a Urgent Recall Notice letter dated January 14, 2011 to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were instructed to discontinue use of Cartridges from the identified lot. All unused cartridges will be credited and are to be returned as instructed on the enclosed Business Reply Card. If you have any questions regarding this information please contact your Abbott Point of Care Technical Support at 800-366-8020 Option 1, or your Abbott Point of Care representative.
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
CHL
Quantity affected
52,750 cartridges
Distribution
USA ( nationwide )
Date initiated
2011-01-01
Date posted
2012-08-02
Date terminated
2013-05-22
Location
Princeton, NJ