i-STAT CREA Cartridge, Abbott List Number: 06F10-01, 02; IVD. Manufactured by Abbott Point of Care, Inc., Princeton, NJ.

FDA device recall Z-0456-2011 · posted 2010-11-24 · Terminated

Recall details

Recalling firm
Abbott Point of Care Inc.
Reason for recall
The i-STAT CREA Cartridges may produce inaccurate (depressed) results at or near the end of the labeled expiry date when stored above 25 degrees for 2 weeks.
Action taken
Abbott Point of Care issued Urgent Product Recall Notification letters dated March 2010 to all direct customers identifying the affected product and the actions to be taken. Customers were instructed to determine if they had cartirdges that had been stored above 25 degrees Celsius, and if so, contact Abbott for replacement or credit. Abbott Technical Support can be contacted at 800-366-8020 option 1 about this recall.
Root cause
Other
Status
Terminated
Product code
CGL
Quantity affected
34525 cartridges
Distribution
Worldwide Distribution: Throughout USA and Canada.
Date initiated
2010-03-17
Date posted
2010-11-24
Date terminated
2011-09-22
Location
Princeton, NJ