Hydration Station with Pedestal (Part # E30003-01), Hydration Station with Pedestal & DVD (Part # E30003-01D), Sybaritic

FDA device recall Z-1245-2009 · posted 2009-07-15 · Terminated

Recall details

Recalling firm
Sybaritic, Inc
Reason for recall
Sybaritic conducted a recall on the Hydration Station 806 because of the potential for hot water to spurt through steam ports into unit and cause burns. Also, potential for customers to inadvertently place their legs over the side wall of the unit directly onto the steam ports and cause burns. There were 18 injuries and 8 malfunctions reported.
Action taken
Sybaritic, Inc. issued a letter dated May 15, 2007 addressed to "Hydration Station Customer". The letter informed the customer of the new 2007 Hydration Station Upgrade Improvement Kits including instructions and advised the customer to coordinate the installation of the components with their Technical Service Department. Further questions can be addressed to Sybaritic, Inc. Technical Administration at 1-800-445-8418 extension 373.
Root cause
Device Design
Status
Terminated
Product code
IMB
Quantity affected
370
Distribution
Worldwide Distribution - US including states of AL, AZ, CA, CO, CT, FL, GA, ID, IL, IN, IA, KS, KY, LA, MD, MA, MI, MN, MS, MO, MI, NE, NV, NH, NJ, NM, NY, NC, OH, OR, PA, SC, TN, TX, UT, VA, WA, WV a
Date initiated
2007-05-15
Date posted
2009-07-15
Date terminated
2012-07-05
Location
Bloomington, MN