Hydradjust IV DR Urological Table Sedecal Generator (p/n 750732 - 64W generator & p/n 750733 80W generator). Mallinckrod

FDA device recall Z-1904-2010 · posted 2010-06-25 · Terminated

Recall details

Recalling firm
Mallinckrodt Inc
Reason for recall
Covidien received complaints of the Sedecal Generator failing and producing smoke at customer sites. The failure modes are located in different parts of the generator and will lead to the generator becoming inoperative.
Action taken
Covidien Issued an "Urgent Device Correction" notification dated June 14, 2010 informing their customers of the potential problem. Consignees were instructed that Field Service Engineers will be in contact arrange a service visit to perform corrections on the affected units. For further information, contact Covidien Product Monitoring at 1-800-778-7898.
Root cause
Process control
Status
Terminated
Product code
IZO
Quantity affected
144 units
Distribution
Worldwide Distribution -- United States, Canada and China.
Date initiated
2010-06-14
Date posted
2010-06-25
Date terminated
2012-02-25