Atlas FDA

human med Biofill Infiltration Cannula, diameter 2.5mm, 15cm, REF 2024-012, REF 1580107, sterile, single use.

FDA device recall Z-1251-2025 · posted 2025-02-28 · Open, Classified

Recall details

Recalling firm
HUMAN MED AG
Reason for recall
The cannula tubes do not comply with the specification. The water jet emerging from the nozzle of the nonconforming cannulas may emerge as a spot jet rather than a fan-shaped spray jet.
Action taken
The recalling firm issued a Field Safety Notice to Distributors dated 10/18/2025 via email explaining the reason for the Field Safety Notice, the potential hazard, and informing the consignee to discontinue delivering the affected products to customers and users and to dispose of the product if still in their warehouse. A Field Safety Notice was enclosed to be issued to the consignee's users and customers who received the affected product from the Distributor. The User and Customer letter dated 10/17/2024 was similar to the Distributor letter except provided instructions to check the spray pattern as instructed in the user manuals prior to using the cannula. If the pattern was not in a flat, fan-shaped pattern, they are not to use the cannula and it should be sorted out and disposed of properly. A confirmation form was enclosed to confirm the cannula has been correctly disposed of and the form was to be sent to the distributor or to the email address provided.
Root cause
Process control
Status
Open, Classified
Product code
QPB
Quantity affected
775 devices
Distribution
US Nationwide distribution in the state of FL.
Date initiated
2024-10-18
Date posted
2025-02-28
Location
Schwerin, Germany

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
BODY-JET (K082025)Human Med AG2008-08-13