Human lgG Subclass Liquid Reagent Kits for use on the Siemens BN"IJ Analyser Product Code: NK001.T This kit is intended

FDA device recall Z-1072-2017 · posted 2017-01-23 · Terminated

Recall details

Recalling firm
The Binding Site Group, Ltd.
Reason for recall
Customer complaints identified that samples containing extremely high IgG4 levels exhibit antigen excess and will yield spuriously low IgG4 results.
Action taken
Binding Site sent an Important Information letter dated September 30, 2014, Customers were instructed to retain the notification for records and additional instructions with the firm stating they were updating the instructions for use. Customers with questions were advised to contact their Binding Site representative or technical.service@bindingsite.co.uk.
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
CFN
Distribution
Worldwide Distribution - US including n FL, OH, TN, MN, NW, MI, CA. and Internationally to Canada, France, UK, Spain, Germany, Czech Republic, Belgium, Italy, Australia, Brazil, Chile, China, Costa Ri
Date initiated
2014-09-11
Date posted
2017-01-23
Date terminated
2017-05-03
Location
Birmingham, United Kingdom