Atlas FDA

HUDSON RCI mBrace, ET Tube Holder with Bite Block and 2 Point Head Strap, REF DYNJMBRC2B; tracheal tube fixation device

FDA device recall Z-0696-2025 · posted 2024-12-18 · Open, Classified

Recall details

Recalling firm
MEDLINE INDUSTRIES, LP - Northfield
Reason for recall
Complaint defects have been noted regarding the adhesive elements of the device and the frame of the holder putting additional pressure on the patient's cheekbones and upper lip which may cause dermal injury depending on the device positioning.
Action taken
Medline Industries issued an URGENT MEDICAL DEVICE RECALL to its consignees of 11/18/2024 via USPS and email. The notice explained the issue, risk to user, and requested the destruction of the affected product. For questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.
Root cause
Nonconforming Material/Component
Status
Open, Classified
Product code
CBH
Quantity affected
360 uinits
Distribution
US and Canada
Date initiated
2024-11-18
Date posted
2024-12-18
Location
Northfield, IL

Monitor recalls programmatically