Hudson RCI Concha Therm Neptune Humidifier, Rx Only, Product Usage: Usage: Device that is intended to add moisture to, a
FDA device recall Z-2036-2012 · posted 2012-07-19 · Terminated
Recall details
- Recalling firm
- Teleflex Medical
- Reason for recall
- Speaker Failure. If the speaker fails and an event occurs which would generate an audio alarm, the alarm will not sound and could cause a potential delay in patient treatment.
- Action taken
- Teleflex sent Urgent Medical Device Notification letters dated June 12, 2012 to all affected customers. The letter identified the affected products, problem and actions to be taken. Customers were instructed to complete the enclosed Acknowledgement Form and return per the instructions provided. Once the form is received, Teleflex Medial will schedule for a technician to come to their facility and exchange the affected product. For questions call your local sales representative or Customer Service at 1-866-246-6990.
- Root cause
- Device Design
- Status
- Terminated
- Product code
- BTT
- Quantity affected
- 14,794 each
- Distribution
- Worldwide Distribution - U.S. (nationwide) and the countries of Argentina, Australia, Brazil, Canada, Chile, China, Europe, India, Japan, Mexico, South Korea, Taiwan and UK.
- Date initiated
- 2012-06-12
- Date posted
- 2012-07-19
- Date terminated
- 2014-01-28
- Location
- Research Triangle Park, NC