HTA Console System, 110 Volt, Product Number: 56000 (HTA system is a software-controlled hysteroscopic thermal endometri

FDA device recall Z-0687-2007 · posted 2007-03-31 · Terminated

Recall details

Recalling firm
Boston Scientific Corporation
Reason for recall
Defective integrated circuit board could result in loss of the system pump and patient injury (hot fluid 90 degree C into uterus) if in use during the recirculation/heating phase of treatment
Action taken
Boston Scientific notiifed accounts by letter dated January 24, 2007 via federal Express Mail. Accounts were requested to examine inventory, remove and return product.
Root cause
Other
Status
Terminated
Product code
MNB
Quantity affected
57 units
Distribution
Worldwide-USA, UK, Denmark, Italy, and France
Date initiated
2007-01-24
Date posted
2007-03-31
Date terminated
2007-12-12
Location
Natick, MA