Horizon Cardiology ECG Management This is a software application designed for use in various hospital departments to imp

FDA device recall Z-0900-2011 · posted 2011-01-13 · Terminated

Recall details

Recalling firm
McKesson Provider Technologies - Medical Imaging Group
Reason for recall
When importing ECG files from Mortara ECG carts using DICOM interface, the Horizon Cardiology ECG Management System does not display the amplitudes of the waveform properly.
Action taken
McKesson Medical sent an advisory notice letter dated June 16, 2010, to all affected customers. Advisory notice ADV01646 was distributed to 13 affected clinical sites from August 30, 2010 to September 28, 2010. The letters identified the product that is affected, the circumstances under which patient safety issue may occur, the risk, and the immediate actions that need to be taken along with a temporary fix. Customers were instructed to use the ECG cart printout to verify results of procedures acquired by Mortara ECG carts. Users should be aware of the problem when referrring to past ECG procedures and when performing a visual comparison with ECG procedures acquired by Mortara ECG carts. Customers should disable the Glasgow Serial Comparison for ECG procedures. McKesson Support would contact the customer to coordinate the deployment of the software update on all the affected applications and Horizon Cardiology client applications. For any questions regarding this recall, customers should call the Customer Support department at 1-866-777-0202 and speak to their Support Manager.
Root cause
Device Design
Status
Terminated
Product code
DQK
Quantity affected
16 units
Distribution
Nationwide Distribution including AZ, CA, FL, GA, MA, PA, UT, and WI.
Date initiated
2010-06-16
Date posted
2011-01-13
Date terminated
2012-08-17
Location
Parsippany, NJ