Atlas FDA

HORIBA ABX PENTRA 400 I.S.E. Module, Model Numbers: P400ISE110EN02 and P400ISE110US02, All Serial Numbers, Software Vers

FDA device recall Z-1141-2013 · posted 2013-04-18 · Terminated

Recall details

Recalling firm
Horiba Instruments, Inc dba Horiba Medical
Reason for recall
HORIBA Medical is informing all ABX PENTRA 400 Analyzer with I.S.E. Module customers that there are no Linearity Limit Flags for the I.S.E. assays currently programmed into the ABX PENTRA 400 Analyzer software. All other assays run on this analyzer do have Linearity Limit Flags to alert the user to values that fall below or above the validated linearity range for each assay.
Action taken
A recall letter, Field Safety Notice & Acknowledgement mailed to the customers March 20, 2013. The letter included the product information, reason for recall, instructions on what to do with the recalled product, and contact information:"If you have any questions or concerns, please contact your local HORIBA Medical representative. We sincerely apologize for any inconvenience this may cause your laboratory. Please contact your Customer Service Representative at 1-888-903-5001 (option 3) for any additional questions related to this customer letter."
Root cause
Software design
Status
Terminated
Product code
JGS
Quantity affected
905 units
Distribution
Nationwide Distribution.
Date initiated
2013-03-20
Date posted
2013-04-18
Date terminated
2014-09-26
Location
Irvine, CA