Atlas FDA

Homogeneous Enzymic Homocysteine Reagent Two Part Reagent Method. Catalog Number E02-057-02. A homocysteine test system

FDA device recall Z-0921-2012 · posted 2012-01-27 · Terminated

Recall details

Recalling firm
Catch Incorporated
Reason for recall
Homogeneous Enzymic Homocysteine Reagent Two Part Reagent Method will not calibrate due to Reagent R1 had prematurely decayed.
Action taken
Catch Inc. sent an "URGENT-PRODUCT RECALL" letter dated 27 June 2011 to one sole customer. The letter described the product, problem, and actions to be taken by the customer. A reply form was attached to the letter for the customer to complete and return via fax at 425-402-8954. Please contact the firm at 425-402-8960 for questions regarding this recall.
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
LPS
Quantity affected
47 kits
Distribution
Product distributed in PA.
Date initiated
2011-06-27
Date posted
2012-01-27
Date terminated
2012-07-11
Location
Bothell, WA