HomeChoice PRO Automated Peritoneal Dialysis Systems; Manufactured for Baxter Healthcare Corporation, Deerfield, IL 6001
FDA device recall Z-1019-05 · posted 2005-07-19 · Terminated
Recall details
- Recalling firm
- Baxter Healthcare Renal Div
- Reason for recall
- The screws that secure the Power Entry Module (PEM) to the base of the HomeChoice device are not reliably grounded to the device's ground system, posing an elcetric shock hazard.
- Action taken
- Baxter sent Urgent Device Correction letters worldwide to their customers on 6/21/05 and home patients on 6/22/05. The accounts were informed of teh risk of electrical shock from screws used to secure the Power Entry Module to the back of the HomeChoice system, and were instructed to always connect the power cord to the power entry module first, before plugging the power cord into a grounded electrical outlet. The users were advised to use caution to avoid contact the power entry module screws while powering on or powering off the device, and were informed that a new mounting bracket that providews more reliable grounding was installed in instruments processed through the Baxter service organization as of May 2005. All remaining instruments will receive the new mounting bracket by August 2006, and Baxter will contact the accounts to arrange for the upgrade. Any questions were directed to Baxter Global Technical Services at 1-800-553-6898, prompt 1.
- Root cause
- Other
- Status
- Terminated
- Product code
- FKX
- Quantity affected
- 20,133 units
- Distribution
- Nationwide, including Puerto Rico, and internationally to Argentina, Austria, Australia, Belgium, Canada, Chile, China, Colombia, Denmark, Ecuador, France, Germany, Guatemala, Hong Kong, Ireland, Indi
- Date initiated
- 2005-06-21
- Date posted
- 2005-07-19
- Date terminated
- 2007-12-11
- Location
- Park, IL
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|