Atlas FDA

Hologic Discovery series Bone Densitometer System Software versions 12.6, 2.3., or 2.4. Indicated for the estimation of

FDA device recall Z-1725-2009 · posted 2009-08-25 · Terminated

Recall details

Recalling firm
Hologic, Inc.
Reason for recall
System may not turn off at the end of the x-ray and cause additional x-ray exposure
Action taken
Hologic issued an Urgent: Medical Device Recall customer notification letter, dated April 17, 2009, by certified mail to inform customers of the problem. With the notification letter, customers will receive the following: -Software version Apex 12.6.2, 2.3.2, or 2.4.2 -Instructions to upgrade their systems with the provided software and to verify the new version on the system as a confirmation that the software was successfully loaded. -New Operator's Manual for installations where the operating characteristics of the system have changed due to the software upgrade. For Customers with systems with software versions other than 12.6, 2.3, or 2.4: Customers were asked to upgrade their QDR software, return their old QDR software disk, and complete and return to Hologic the enclosed Confirmation of Notification.
Root cause
Component design/selection
Status
Terminated
Product code
KGI
Quantity affected
1914 units total
Distribution
Worldwide Distribution -- USA, Canada, Europe, Austria, Belgium, Bulgaria, Czech Republic, Denmark, France, Germany, Great Britain, Greece, Hungary, Iceland, Italy, Latvia, Lithuania, Macedonia, Nethe
Date initiated
2009-04-07
Date posted
2009-08-25
Date terminated
2010-11-29
Location
Bedford, MA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
APEX 2.0 SOFTWARE FOR QDR X-RAY BONE DENSITOMETERS (K072847)Hologic, Inc.2008-03-28