HLD System, Model 610, Medical Device Cleaning and High Level Disinfection Washer/Pasteurizer.
FDA device recall Z-1995-2014 · posted 2014-07-06 · Terminated
Recall details
- Recalling firm
- Cenorin, LLC
- Reason for recall
- The temperature sensor/control system in the HLD Systems Model 610 may provide a false temperature reading. This caused the water to remain at ambient temperature throughout the pasteurization cycle rather than using hot water.
- Action taken
- Cenorin sent a Notification letter dated December 18, 2013 to all affected customers. The letter identified the affected product, problem and actions to be taken. The letter stated: Cenorin has developed an upgrade to mitigate this issue as well as provide additional process control and system monitoring benefits. Cenorin included a Return Response form with the recall letter to be completed and return. For questions contact HLD Systems Sales Representative for questions.
- Root cause
- Device Design
- Status
- Terminated
- Product code
- LDS
- Quantity affected
- 25 HLD systems
- Distribution
- Worldwide Distribution - US (nationwide): CA, DE, FL, MA, MI, NY, PA, TX, UT, WA, and WI and the countries of Uruguay and Taiwan.
- Date initiated
- 2013-01-24
- Date posted
- 2014-07-06
- Date terminated
- 2015-02-10
- Location
- Kent, WA