Hitachi Echelon Magnetic Resonance Imaging System, Serial Number: V-001, equipped with CTL coil type: MR-CTL-150.

FDA device recall Z-0152-2007 · posted 2006-11-04 · Terminated

Recall details

Recalling firm
Hitachi Medical Systems America Inc
Reason for recall
Component defect/overheating device; A malfunctioning decoupling circuit in the CTL coil caused high current flow and circuit overheating.
Action taken
The recalling firm issued a Device Correction letter to the clinic customer on 9/15/2006 making them aware of this potential problem and of their completed corrective action on 8/14/2006, by a Service Rep visit.
Root cause
Other
Status
Terminated
Product code
MOS
Quantity affected
1 unit
Distribution
The recalled unit was distributed to a single customer.
Date initiated
2006-08-14
Date posted
2006-11-04
Date terminated
2012-09-19
Location
Twinsburg, OH