Hitachi CXR4 Computed Tomography (CT) Scanners, Software Version 1.51 or less.

FDA device recall Z-1651-2009 · posted 2009-07-07 · Terminated

Recall details

Recalling firm
Hitachi Medical Systems America Inc
Reason for recall
Incorrect scale on image: A software error can occur if two (2) different Field-of-View (FOV) settings are registered in a protocol using Multi-Reconstruction. This could cause the image generated to be displayed with the wrong scale factor rendering anatomical measurements incorrect.
Action taken
HMSA issued an URGENT - DEVICE CORRECTION notification to their customers on 4/14/2009. The notification describes the potential software error and the condition under which these image distortions can occur. The notification goes on to provide the customer with a work-around procedure which they can use to avoid encountering this issue with their current software configuration(s). The letter states that the firm will be installing updated software on all affected systems beginning in June, 2009, and that a Hitachi Service Rep. will be contacting each customer and making arrangements to come out and install a software upgrade on the affected system(s) at no cost to the customer. Lastly, the letter provides a telephone # to the firm's Applications Helpline: 1-877-228-2777, which the customer can use to obtain additional information, if desired.
Root cause
Software design
Status
Terminated
Product code
JAK
Quantity affected
47 units
Distribution
Nationwide distribution, including Puerto Rico (PR).
Date initiated
2009-04-14
Date posted
2009-07-07
Date terminated
2012-09-21
Location
Twinsburg, OH