Atlas FDA

Hitachi Altaire Magnetic Resonance Imaging System

FDA device recall Z-1103-06 · posted 2006-06-10 · Terminated

Recall details

Recalling firm
Hitachi Medical Systems America Inc
Reason for recall
Circuit Failure-There is the potential risk of the Phase Loss Detector Safety Device (PHSEN PCB) malfunctioning in the event of a power loss. The circuit may fail to turn off the system main breaker when 2 of the 3 power phases are lost.
Action taken
The recalling firm sent a Device Correction Letter dated 1/27/06 to all of their customers. The letter informs the customers of the problem in the (PHSEN PCB) and notifies the customers that an HMSA Field Service Technician will be performing a PCB modification to the unit, on-site, during their next scheduled preventative maintenance visit. Should the customer wish to have the modification performed prior to the representative''s scheduled visit, the letter says that they can telephone their local Field Service Rep. and make arrangements to have the modification performed at their convenience.
Root cause
Other
Status
Terminated
Product code
LNH
Quantity affected
213 devices have been distributed and installed.
Distribution
Nationwide-The device was distributed to consignees located throughout the United States.
Date initiated
2006-01-27
Date posted
2006-06-10
Date terminated
2008-11-13
Location
Twinsburg, OH

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
HITACHI ALTAIRE (K022390)Hitachi Medical Systems America, Inc.2002-10-11
HITACHI ALTAIRE (K002420)Hitachi Medical Systems America, Inc.2000-10-17