HiResolution" Bionic Ear System - M Battery Charger Model Number CI-5607
FDA device recall Z-1813-2025 · posted 2025-05-22 · Open, Classified
Recall details
- Recalling firm
- Advanced Bionics, LLC
- Reason for recall
- Due to incorrect the battery charger being packaged and distributed. Product labeling does not match the intended product as specified.
- Action taken
- On 03/06/2025, customer service informed customers via telephone call of a distribution error. An updated "URGENT: Medical Device Recall Notification will be communicated to customers on or about 04/09/2025, via email. This updated notification will inform them that upon further review/investigation, it was identified that the individual package label was for the Marvel M Battery Charger (CI-5607), however the product within the package was a Naida CI PowerCel Charger (CI-5605). Customer are informed that: 1. The mislabeled battery charger they received was from lot either Lot Numbers: 410488967 and/or 410488966 Customers are instructed to: 2. Return the affected chargers to Advance Bionics. Advance Bionics will provide the correct charger to customer and will confirm receipt and return of the affected chargers. For questions contact customer service at 877-829-0026 or email Customerservice@advancebionics.com
- Root cause
- Under Investigation by firm
- Status
- Open, Classified
- Product code
- MCM
- Quantity affected
- 6 chargers
- Distribution
- U.S.: AZ, FL, MI, SC, TN and TX. O.U.S.: N/A
- Date initiated
- 2025-03-05
- Date posted
- 2025-05-22
- Location
- Valencia, CA