Atlas FDA

HipLOC CHS Plate/Lag Screw Introducer

FDA device recall Z-0885-2016 · posted 2016-02-25 · Terminated

Recall details

Recalling firm
Biomet, Inc.
Reason for recall
A complaint was received where the Hiploc Plate/Lag Screw Introducer was undersized and would not accept the introducor coupling screw. It appears now that not all parts manufactured to Rev A drawing were reworked.
Action taken
Biomet sent an " URGENT MEDICAL DEVICE RECALL NOTICE" dated February 2016, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The following actions are REQUIRED: Immediately locate and remove the identified device(s) listed below from circulation. Remove the affected product from circulation. Review this notification and ensure that all affected personnel are aware of its contents. Carefully follow the instructions on the enclosed Response Form. Email a copy of the response form to CPWARFieldAction@zimmerbiomet.com prior to return of product. You do not need to complete the Biomet UK response form. Use priority carrier for your shipment. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA: "Online: www.fda.gov/Safety/MedWatch/HowToReport/default.htm (form available to fax or mail), or "Call (800)FDA-1088 Thank you in advance for your assistance and prompt attention. On behalf of Zimmer Biomet, I apologize for any inconvenience this may cause. Questions related to this notice should be directed to (574) 372-1570, Monday through Friday, 8 a.m. to 5 p.m.
Root cause
Process change control
Status
Terminated
Product code
LXH
Quantity affected
23 units
Distribution
US Distribution to the states of : MO, UT, AZ, IL, PA, KY, MI, NC, NJ, MN and GA., and Internationally to The Netherlands.
Date initiated
2016-02-16
Date posted
2016-02-25
Date terminated
2017-02-14
Location
Warsaw, IN