Atlas FDA

Hinged Knee Support***LATEX FREE" Product Usage: Provide added lateral stability and allows the knee to tolerate more st

FDA device recall Z-1605-2014 · posted 2014-05-10 · Terminated

Recall details

Recalling firm
Zimmer, Inc.
Reason for recall
During a transfer of products from a recently shutdown facility, the firm discovered raw material labeled as latex free actually contained latex.
Action taken
Zimmer sent an Urgent: Device Removal Letters dated April 8, 2014 to their customers. The letter identified the affected product, probllem and actions to be taken. Customers were instructed to review this notification, iIdentify and quarantine the affected product to prevent further distribution or use, complete the attached Response Form and return it via email. orporateQuality_PostMarket@zimmer.com. Return affected product to: Zimmer Surgical Attn: QA/RA Dept.- Recall 200 West Ohio Avenue Dover, Ohio 44622 USA Please include a copy of the Response Form with the shipment. For returns outside the US, please email Rhonda.duncan@zimmer.com obtain an IRA (international return authorization) number. The IRA request should include the part number(s) being returned and the quantity. Please write the associated IRA number on the outside of the box. 4. Zimmer will credit your account for returned Drop-Lok" Knee Braces, Cartilage Knee Braces, Hinged Knee Supports, or Neoprene Tennis Elbow Supports. Please return a copy of the completed response form along with your returned product to ensure proper credit. Important: Please distribute this notification to all personnel within your organization who need to be aware. If you have further transferred affected product(s), please provide the customer's information on the Business Response Form to Zimmer. For questions call 330-354-0989
Root cause
Labeling mix-ups
Status
Terminated
Product code
DWL
Quantity affected
546 units
Distribution
Worldwide Distribution - US Nationwide in the states of: AK, AZ, CA, FL, GA, IA, IL, IN, LA, MD, MI, MO NC, NE, NV, NY, OH, OK, OR, PA, RI, SD, TX, UT, VA, VT, WA, WI & WV and countries of: AUSTRALIA,
Date initiated
2014-03-19
Date posted
2014-05-10
Date terminated
2015-04-14
Location
Warsaw, IN

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.