Atlas FDA

Hillrom Temperature Management Controller, MP, Model # 2083516; used in conjunction with Warming Blankets and Mattresses

FDA device recall Z-0841-2025 · posted 2025-01-10 · Completed

Recall details

Recalling firm
Augustine Temperature Management, LLC
Reason for recall
HotDog Temperature Management Controllers Models WC71, WC77, and Temperature Management Controller MP Model 2083516 may experience a Power-on self-test error code "EA POST".
Action taken
Augustine Temperature Management, LLC began correcting (installing ferrites and software updates) the affected devices February, 2022 and completed the actions in August 2023. The firm did not perform any notification to its consignees that corrections would be implemented.
Root cause
Device Design
Status
Completed
Product code
DWJ
Quantity affected
77 units
Distribution
Worldwide
Date initiated
2022-03-14
Date posted
2025-01-10
Location
Minnetonka, MN